6-7-1
Performing the Battery Calibration Procedure
NOTE:
It is recommended to commence this procedure towards the end of the working day (when the
battery is fully charged) then allow the procedure to be completed overnight, as outlined in
Figure
1) Click on the battery icon at the bottom of the scanning screen
OR
from the message window shown below, click Recalibrate.
The Battery Calibration Wizard (Page 1) opens -
2.) Make sure the AC power cable is connected to the Vivid i n, then click Next to commence the first
stage of the process (Full Charge to maximum capacity).
This step will fully recharge the battery and may take up to 4 hours.
6-8
5
EVISION
6-6.
Figure 6-7 Battery Calibration Wizard - Page 1
Section 6-7 - Battery Calibration
V
Figure
6-7.
S
M
IVID I N
ERVICE
ANUAL