Accessing The Diagnostic Test Window; Diagnostic Test Window - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
7-2-3

Accessing the Diagnostic Test Window

The diagnostic tools are accessed from 2D-Mode by simultaneously pressing <ALT+Config> on the
alphanumeric keyboard. The Diagnostic Test window is displayed, as shown below:
Diagnostic
Test
Tree
Current Test
Controls &
Special
Instructions
Status Area
Note:
For a summary of the various Diagnostic Test Menu options, together with details of the
information displayed in different areas of the screen, see
Note:
When performing a specific test for an intermittent problem, it is highly recommended to use the
Loop option (use the combo box or type the required number of loops) as appropriate. Looping
the test will assist in detecting the problem.
Note:
Once the test has been completed, a log can be viewed, as described in the 7-2-3-1.
5
EVISION
Figure 7-1 Diagnostic Test Window
Chapter 7 - Diagnostics/Troubleshooting
V
IVID I N
Table 7-2
on page 7-4 and 7-2-4.
S
M
ERVICE
ANUAL
Data Flow
Map
7-3

Advertisement