Gehc Current Leakage Limits; Chassis Current Leakage Limits - Accessible Metal Surfaces - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
10-7-2

GEHC Current Leakage Limits

The following limits are summarized for NFPA 99 (for USA) and IEC 60601-1 Medical Equipment Safety
Standards. These limits are GEMS standards and in some cases are lower than the above standards
listed.
Table 10-13 Chassis Current Leakage Limits - Accessible Metal Surfaces
Country
Table 10-14 Type BF Applied Part Current Leakage Limits
Country
USA
Other
a. Mains Applied is the Sink Leakage Test. In this test, mains (supply) voltage is supplied to the part or equipment to
determine the amount of current that will pass (sink) to Ground if a patient comes into contact with the part.
Table 10-15 Type CF Applied Part Current Leakage Limits - Surgical Probes and ECG Connections
Country
USA
Other
a. Mains Applied is the Sink Leakage Test. In this test, mains (supply) voltage is supplied to the part or equipment to
determine the amount of current that will pass (sink) to Ground if a patient comes into contact with the part.
5
EVISION
Normal Condition
N/A
USA
0.1 mA
Other
- Non-Conductive (Floating) Surface and Cavity Probes
Normal
Condition
0.05 mA
0.1 mA
Normal
Condition
0.01 mA
0.01 mA
Chapter 10 - Care and Maintenance
Open Ground
0.3 mA
0.5 mA
Open
Ground
Reverse Polarity
0.05 mA
0.05 mA
0.5 mA
0.5 mA
Open
Ground
Reverse Polarity
0.05mA
0.05 mA
0.05 mA
0.05 mA
V
S
IVID I N
ERVICE
Reverse Polarity
Open Neutral
0.3 mA
0.5 mA
Open
Neutral
Mains Applied
0.05 mA
0.5 mA
Open
Neutral
N/A
0.05 mA
M
ANUAL
N/A
0.5 mA
a
N/A
5.0 mA
Mains
a
Applied
0.025 mA
0.05 mA
10-21