Set Up Connection To A Dicom Server In A Network; Dicom Server Ip Address Setting On The Scanner - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
3-8-4

Set Up Connection to a DICOM Server in a Network

In this case, the Vivid i n is configured to work with a DICOM server in a network environment. Images
are first saved on the local image buffer on the scanner. At the end of the examination, the images are
sent to the DICOM server via a DICOM spooler.
This scenario requires that the scanner is configured to be connected to the DICOM server as described
below.
3-8-4-1
Overview
To work against the DICOM server, the following information has to be entered in the scanner:
DICOM server IP address, subnet mask and eventually the gateway.
DICOM server port number.
DICOM server AE title (the server application's name).
Name of Device.
3-8-4-2

DICOM Server IP Address Setting on the Scanner

1.) Press CONFIG (F2) and log on as Adm, as described in "Log On to the System as ADM" on page 4-
10.
2.) Select CONNECTIVITY (in the lower part of the window).
3.) Select the DATAFLOW tab.
4.) Select the arrow to the right of the Name field to list all dataflows in a pull-down menu, see
Figure
5.) From the Dataflow Name pull-down list, select the dataflow you want to configure - see
5
EVISION
3-70.
Figure 3-70 Select Dataflow
Chapter 3 - System Setup
V
S
M
IVID I N
ERVICE
ANUAL
Arrow
Select dataflow from
Dataflow Name pull-down
menu.
Figure
3-70.
3-81