GE H
EALTHCARE
D
FL091021, R
IRECTION
8-3-7
LCD Display Replacement Procedure
8-3-7-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the LCD Display replacement
procedure.
8-3-7-2
Time Required
15 minutes
8-3-7-3
Preparation
Shut down the Vivid i n ultrasound unit, as described in
8-3-7-4
LCD Display Removal Procedure
1) Remove the Control Panel and Keyboard as described in the
Removal Procedure" on page
2.) Remove the LCD Display Frame as described in the
page
3.) Loosen and remove the four screws that fasten the LCD Display to the Rear Cover & Latch
Assembly, as shown in shown in
5
EVISION
8-22.
Figure 8-43 Removing the Screws from the LCD Display
Chapter 8 - Replacement Procedures
Chapter
8-6.
"LCD Display Frame Removal Procedure" on
Figure 8-43
below.
V
S
IVID I N
ERVICE
3-
System
Setup.
"Control Panel and Keyboard
M
ANUAL
8-45