Control Panel and Keyboard Replacement Procedure

8-2-3-1

Tools

Use the appropriate Phillips and flat screw drivers, as indicated in the following Control Panel and
Keyboard (and sub-component) replacement procedures.
8-2-3-2
Time Required
10 minutes
8-2-3-3
Preparations
Shut down the Vivid i n ultrasound unit, as described in
8-2-3-4

Control Panel and Keyboard Removal Procedure

1) Place the Vivid i n face down and remove the battery as described in the
Procedure" on page
2) Loosen and remove each of the 9 screws that fasten the Control Panel and Keyboard onto the
Bottom Assembly, as indicated in
Figure 8-6 Removing the Fastening Screws from the Bottom Assembly
3.) Turn the Vivid i n over, release the front latches and open the LCD Display cover.
8-6
5
EVISION
8-2.
Section 8-2 - External Component Replacement Procedures
Chapter
Figure 8-6
below.
V
S
IVID I N
ERVICE
3-
System
Setup.
"Battery Removal
M
ANUAL
This screw
not captive
(do not
lose!)