Configuring the Vivid i n for eVue Dataflow

1.) Verify which PC in the hospital/institute has been designated as the e-Vue station.
2.) Make sure the Vivid i n is connected to the Network via cable or wireless.
3.) On the Vivid i n scanner, press the Config button.
The System Configuration window opens.
4.) Click Connectivity.
The Connectivity dialog box opens.
5.) Select the Dataflow Tab, as shown in Figure 3-95 on page 3-105 .
6.) Click on the Name field to access the drop-down menu.
7.) From drop-down menu, select the Dataflow named Local Archive – Int. HD/eVue.
8.) Trackball to the Selected devices window and select (highlight) eVue (see
5
EVISION
Figure 3-95 Connectivity - Dataflow Tab Selected
Chapter 3 - System Setup
V
S
M
IVID I N
ERVICE
ANUAL
Figure
3-95).
3-105