Service Adjustments; Lcd Display Adjustments And Calibration - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
Chapter 6

Service Adjustments

Section 6-1
Overview
6-1-1
Purpose of Chapter 6
This section provides references to the procedures required for testing and adjusting the Vivid i n.
These tests are optional, and may be used to check the system for errors.
NOTE:
There are no voltage adjustments required.
Section 6-2

LCD Display Adjustments and Calibration

Overall low image quality or "shades" visible on the Vivid i n LCD Display may be the result of improper
monitor calibration. Therefore, adjustments may be made to the Contrast, Brightness, Watermark and/
or Blue Tint, as required.
LCD Display adjustments should be made in the correct order, as stated in the following procedures.
NOTICE
NOTE:
Incorrect calibration of the Vivid i n monitor may also result in less than optimal image quality
when reviewing images on a DICOM or EchoPAC station.
5
EVISION
Chapter 6 - Service Adjustments
V
S
M
IVID I N
ERVICE
ANUAL
6-1