GE
D
KTI106052, R
IRECTION
5-13-2-6
Delete all Patients
1.) Click the DELETE ALL PATIENTS... button. Following WARNING message appears on the screen.
WARNING
WARNING
If you select the YES button, all patients data, studies, images and measure report data
will be deleted permanently from the hard disk and cannot be recovered!
!! WARNING:
5-13-2-7
Export System Data
Select the EXPORT SYSTEM DATA button on the "Service Tools" page to Full Backup the System
State. This includes dump-files and text files, the full Service Database informations about probes,
boards, Software, Options and the Event Log File to the DVD Drive.
Operation see:
5-13-2-8
Battery Calibration
Select the BATTERY CALIBRATION button to perform fuel gauge recalibration of the battery.
Operation see:
5-13-2-9
Keyboard Layout
To change the keyboard layout to different languages.
Operation see:
NOTE:
Reconfigure the layout of the keyboard is only useful by changing the concerned keys also.
see:
Section 8-7 "Replacement of Key Caps (by special native language keys)" on page
5-13-2-10
BootUp Screen
With this button you can give in your own name. You can see it, when you Bootup the Voluson i /
Voluson e.
5-13-2-11
Monitor Test
Select the MONITOR TEST button to perform color calibration.
Operation see:
5-13-2-12
Printer
Installation of printers or editing their settings is possible without entering the Windows Desktop.
Operation see:
Only accessories explicitly recognized and released by the system manufacturer GE
WARNING
WARNING
may be used in connection with the system.
!! WARNING:
5-13-2-13
Probe EEPROM
NOTE:
"Probe EEPROM" is available from software 8.2.2 and software CC200 BT'14 and higher.
It is recommended to connect and re-program one probe after the other.
NOTICE
10
EVISION
Figure 5-30 Warning message
Section 7-4-2 "Export Log's and System Data" on page
3-6-4-2-1 "Recalibration by means of Internal Battery Charger" on page
Section 6-9 "Modification of Keyboard Layout" on page
Section 6-3 "LCD Display (Monitor) Adjustment" on page
3-7-4 "Printer Installation manually" on page
Chapter 5 - Components and Functions (Theory)
V
OLUSON
7-9.
6-10.
6-2.
3-87.
i / V
e
OLUSON
S
M
ERVICE
ANUAL
3-77.
8-25.
5-55