Select SpO2's color
Config settings
Data update period
Sensor Cleaning and Disinfection
Procedures
Calibration
12-10
NELLCOR® OXIMAX® SpO2: Sensor Cleaning and Disinfection
This option lets you change the color of this parameter's vital signs that appear
on the Monitor screen.
This option allows you to view, but not change, the various configuration
options. To change any config option, you must go into the password-protected
Configuration Mode. Refer to the Dash 2500 Patient Monitor Service Manual
(PN 2042481-001) for more information.
During normal measurement conditions in the Normal response mode, the
averaging time is 6 to 7 seconds.and the response time is 2 to 4 seconds in Fast
mode. The sum of alarm condition delay and alarm signal generation delay less
than 1 second.
Adhesive sensors are sterile and for single use only. Reusable sensors should be
cleaned before reuse with a 70% alcohol solution. If low-level disinfection is
required, use a 1:10 bleach solution. Do not use undiluted bleach (5% - 5.25%
sodium chlorite) or any cleaning solution other than those recommended here
because permanent damage to the sensor could occur. Do not sterilize the
sensor by irradiation, steam, or ethylene oxide. If disposable sensors or their
packaging are damaged, they must be disposed.
1.
Saturate a clean, dry gauze pad with the cleaning solution. Wipe all surfaces
of the sensor and cable with this gauze pad.
2.
Saturate another clean, dry gauze pad with sterile or distilled water. Wipe all
surfaces of the sensor and cable with this gauze pad.
3.
Dry the sensor and cable by wiping all surfaces with a clean, dry gauze pad.
The SpO
parameter does not require calibration.
2
Dash 2500 Patient Monitor
2042480-001 C