Peripheral Update For Ec Countries - GE LOGIQ C5 Quick Manual

Hide thumbs Also See for LOGIQ C5:
Table of Contents

Advertisement

LOGIQ C5/C5 PRO Quick Guide

Peripheral Update for EC countries

The following is intended to provide the users in EC countries with updated information concerning the connection of the LOGIQ C5/C5 PRO to image recording and
other devices or communication networks.
Peripherals Used in the Patient Environment
The LOGIQ C5/C5 PRO has been verified for overall safety, compatibility and compliance with the following on-board image recording device:
Sony UP-D897 Digital Printer
Sony UP-D23MD Color Printer
HP470b Officejet Printer
The LOGIQ C5/C5 PRO may also be used safely while connected to devices other than those recommended above if the devices and their specifications,
installation, and interconnection with the system conform to the requirements of IEC/EN 60601-1-1
The connection of equipment or transmission networks other than as specified in the user instructions can result in an electric shock
CAUTION
hazard or equipment malfunction. Substitute or alternate equipment and connections requires verification of compatibility and conformity to
IEC/EN 60601-1-1 by the installer. Equipment modifications and possible resulting malfunctions and electromagnetic interference are the
responsibility of the owner.
General precautions for installing an alternate device in patient environment would include:
1. The added device must have appropriate safety standard conformance and CE Marking.
2. There must be adequate stability of the combination between LOGIQ C5/C5 PRO and Peripherals.
3. Risk and leakage current of the combination must comply with IEC/EN 60601-1.
4. Electromagnetic emissions and immunity of the combination must conform to IEC/EN 60601-1-2.
Direction 5272220-100 Rev. 2
41

Advertisement

Table of Contents
loading

This manual is also suitable for:

Logiq c5 pro

Table of Contents