Technical data
General information
The following text has changed:
All patient-related volume values are normalized to
STAPD conditions.
Ambient conditions
The unit "psi" for ambient pressure has been added.
During operation
Temperature
Ambient pressure
Relative humidity
Height
During storage and transport
Temperature
Ambient pressure
Relative humidity
The conditions for use when using additional devices can limit the environment of use of a system as a
whole. Vaporizers and anesthetic agents can limit the use of an anesthesia workstation with regard to its
temperature range and maximum fresh-gas flow. Therefore when using additional devices, follow the
associated instructions for use.
Supplement Fabius plus XL SW 3.n
All patient-related fresh-gas flow values are
normalized to STPD conditions.
10 to 35 °C (50 to 95 °F)
700 to 1060 cmH
O (hPa) (10.9 to 15.9 psi)
2
20 to 80 % (without condensation)
Up to 3000 m (9843 ft)
-10 to 60 °C (14 to 140 °F)
700 to 1060 cmH
O (hPa) (10.9 to 15.9 psi)
2
10 to 90 % (without condensation)
Technical data
37