Steam sterilization

Use a steam sterilizer that meets the requirements
of the standard ISO 17665. Dräger recommends
steam sterilization with fractionated vacuum.
Prerequisite: The product has been cleaned and
disinfected.
1 Sterilize the product.
2 Check the product for visible damage and
replace if necessary.

Maintenance

Inspection
The following table has changed:
Inspections must be carried out regularly
according to the following guidelines and within the
specified intervals. Technical documentation is
available on request.
Checks
Inspection and safety checks
1) Designation applies to the Federal Republic of Germany; corresponds to the "Recurring safety inspection" in the Federal
Republic of Austria

Safety checks

An additional check point has been added to "Check safety
features".
6 Check safety features:
– Check the functional state of the optical
and acoustic alarm generators.
– Check the functional state of the O
alarm.
– Check the locking device of the anesthetic
vaporizer.
– Check the function of the power failure
alarm and the battery function.
– Check the operation of the battery failure
alarm.
– Check S-ORC functionality.
Supplement Fabius plus XL SW 3.n
Interval
1)
Every 12 months
failure
2
Maintenance
Responsible personnel
Service personnel
35