Adjustments and Correction Constants
Sampler Magnitude and Phase Correction Constants (Test 53)
Determine the Insertion Loss of the Cable at 1 GHz
(8753ES)
1. Press
Preset
Meas
2. Press
Center
1
3. Press
CAL KIT
Cal
.
RESPONSE
4. Connect the 24 inch cable from Port 1 to Port 2, as shown in
Figure 3-3
First Connections for Insertion Loss Measurement (8753ES)
5. Press
and then
THRU
through.
6. Press
Save/Recall
7. Make the connections as shown in
Figure 3-4
Second Connection for Insertion Loss Measurement
3-18
Trans: FWD S21(B/R)
G/n
Span
50
M/µ
CAL KIT: 7mm
DONE: RESPONSE
to save the calibration that you just made.
SAVE STATE
Figure
.
.
RETURN
CALIBRATE MENU
Figure
when the analyzer is done measuring the
3-4.
3-3.
Chapter 3