Perform a system
dye calibration
SeqStudio
Genetic Analyzer Instrument and Software User Guide
™
A system dye calibration requires ~30 minutes to complete.
Prepare the dye set calibration standards and plate as described in the product
information sheet for the dye set.
In the home screen:
1. Touch
Settings4Maintenance and Service4Calibration4Dye Calibration.
2. Touch the injection group for the dye set in the plate, then touch Dye set.
Appendix C Dye calibration and install standard checks
C
Calibrate dyes