Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer User Manual page 219

Hide thumbs Also See for Applied Biosystems SeqStudio Genetic Analyzer:
Table of Contents

Advertisement

Table 17 HID run modules
Run module
HIDAnalysis
[1]
Resolution Range: The range of bases over which the resolution (peak spacing interval divided by the peak width at half-max in a GS600 size
standard sample sized with a third order fit) is ≥1.
[2]
The GeneScan
500 LIZ
Size Standard was not included in the HID validation of the instrument.
SeqStudio
Genetic Analyzer Instrument and Software User Guide
Resolution
Approximate
range
run time
[1]
60–470 bp
39 minutes
Note: The following size standards have not been validated for use with the
instrument. A default size standard definition is not provided in the software.
·
GeneScan
350 ROX
·
GeneScan
400HD ROX
Run modules, read lengths, size ranges, and run times
Sizing precision
60–470: <0.15
Size Standard
dye Size Standard
Appendix E Run modules
Compatible size standards
• GeneScan
500 LIZ
Size
[2]
Standard
• GeneScan
600 LIZ
Size
Standard v2.0
E
219