CAUTION! Cleaning and Decontamination. Use only the cleaning and
decontamination methods specified in the manufacturer's user documentation.
It is the responsibility of the operator (or other responsible person) to ensure
the following requirements are met:
·
No decontamination or cleaning agents are used that could cause a
HAZARD as a result of a reaction with parts of the equipment or with
material contained in the equipment.
·
The instrument is properly decontaminated a) if hazardous material is
spilled onto or into the equipment, and/or b) prior to having the instrument
serviced at your facility or sending the instrument for repair, maintenance,
trade-in, disposal, or termination of a loan (decontamination forms may be
requested from customer service).
·
Before using any cleaning or decontamination methods (except those
recommended by the manufacturer), users should confirm with the
manufacturer that the proposed method will not damage the equipment.
To minimize negative environmental impact from disposal of electronic waste, do not
dispose of electronic waste in unsorted municipal waste. Follow local municipal waste
ordinances for proper disposal provision and contact customer service for information
about responsible disposal options.
WARNING! LASER HAZARD. Under normal operating conditions, the
SeqStudio
Genetic Analyzer is categorized as a Class 1 laser product.
However, removing the protective covers and defeating the interlock(s) may
result in exposure to the internal Class 3B laser. Lasers can burn the retina,
causing permanent blind spots. To ensure safe laser operation:
·
Never look directly into the laser beam.
·
Do not remove safety labels, instrument protective panels, or defeat safety
interlocks.
·
The system must be installed and maintained by a Thermo Fisher Scientific
Technical Representative.
·
Remove jewelry and other items that can reflect a laser beam into your eyes
or those of others
·
Wear proper eye protection and post a laser warning sign at the entrance to
the laboratory if the laser protection is defeated for servicing
·
DO NOT operate the laser when it cannot be cooled by its cooling fan; an
overheated laser can cause severe burns on contact.
H
Appendix H Safety
Instrument safety