11
Chapter 11 Maintain the instrument

Install cathode buffer

Install cathode buffer

SeqStudio

Genetic Analyzer Cathode Buffer Container storage
Condition
Shipping
On-instrument storage
Off-instrument storage
Reuse
Remove the
SeqStudio
Genetic Analyzer
Cathode Buffer
Container
Assemble the
SeqStudio
Genetic Analyzer
Cathode Buffer
Container (CBC)
166
Is shipped at ambient temperature.
Store at 2–8°C upon receipt.
After installation, can be stored for up to 2 weeks when the instrument is powered
on and in Cartridge storage mode.
Before opening, can be stored until expiry date on label at 2–8°C.
Do not remove the CBC from the instrument for storage. Do not reuse.
Perform these steps if you are replacing the Cathode Buffer Container. It is not
necessary to remove the CBC after each run.
1. Touch
, touch
2. Press the release button on the autosampler to open the lid.
3. Lift the CBC out of the autosampler.
Equilibrate the CBC to room temperature (15 minutes to overnight) before
assembling.
On a clean and level surface:
1. Ensure that the buffer is above the fill line.
2. Carefully peel off the seal.
3. Wipe off any buffer on top of the CBC with a lint-free tissue. Ensure that the top
of the container is dry.
Description
Eject plate, then open the instrument door when prompted.
SeqStudio
Genetic Analyzer Instrument and Software User Guide