4
Chapter 4 Use the instrument with the Security, Auditing, and E‑signature (SAE) v2.0 module
E-signature requirements when you start a run
62
The Object Name column lists the items that require signatures:
If the Object Name
lists
A dye set name
An install standard
name
A plate name
The example below indicates that E-signature is configured to require 4 signatures
when you start a run, 1 signature is complete and 3 signatures are pending:
• 2 administrator signatures for dye calibration (dye set name is listed two times):
– 1 signature that specifies the "Reviewed and approved" E-signature meaning
(complete, because 1 signature is required [Signatures Required=1] and 1
signature has been logged [Current=1]).
– 1 signature that specifies the "Checked consumables" meaning (incomplete).
• 1 administrator signature for the plate setup that specifies the "Reviewed and
approved" E-signature meaning (incomplete).
• 1 administrator signature for install run that specifies the "Reviewed and
approved" E-signature meaning (incomplete).
The object is
The most recent dye calibration
for the dye set specified in the
plate setup.
The most recent install run for
the application specified in the
plate setup.
Plate setup
SeqStudio
Genetic Analyzer Instrument and Software User Guide
For information, see
"Sign a dye calibration or an
install run" on page 64
"Sign a dye calibration or an
install run" on page 64
"Sign a plate setup" on
page 63