Chapter 10 Manage the software (instrument)
Manage run settings (instrument)
Manage size
standard
definitions
Manage analysis
settings
142
This function manages the list of size standard definitions that you can select from
during plate set up.
To assign a size standard definition to a plate, see "Assign wells: run module, size
standard, dye set, and kit" on page 84.
A size standard defines the sizes in basepairs of known fragments. It is used to
generate a standard curve. The standard curve is used to determine the sizing of
fragments in unknown samples.
Factory-installed items cannot be edited or deleted. To create a new item from a
factory-installed item, copy, edit, then save the new item.
Size standard definitions are accessible to all users.
1. Access the Manage size standards screen:
From
Plate properties tab
Home screen
2. To create a new size standard:
a. Touch a default size standard or a user-created size standard, then touch
Copy.
b. Enter a name and select a dye color.
c. As needed, touch any of the following, then touch Done:
Note: To change an existing value, add a new value, then delete the original
value.
• One or more fragment size values, then touch Delete.
• Add to add a value.
3. As needed, touch a size standard of interest, then touch Edit or View.
Edit fragment/HID analysis settings
Factory-installed items cannot be edited or deleted. To create a new item from a
factory-installed item, copy, edit, then save the new item.
Analysis settings are accessible only to the user who creates them.
1. Access the Analysis settings screen:
From
Plate properties tab Touch More options4Analysis settings, then select the
Home screen
2. To create new analysis settings:
a. Touch the default analysis settings or user-created analysis settings, then
touch Copy.
Touch More options4Manage size standards.
Touch
Settings4Run settings 4Size standards.
settings to assign to the plate setup.
Touch
Settings4Run settings 4Analysis settings, then
manage the list of settings that you can select from during
plate set up.
SeqStudio
Genetic Analyzer Instrument and Software User Guide