Appendix C Dye calibration and install standard checks
Calibrate dyes
194
1. Touch
Settings4Maintenance and Service4Calibration4Calibration
history.
2. Examine the entry for the dye set of interest. If the dye set does not list a date and
a cartridge serial number for a manual calibration or a date for auto calibration,
the dye set requires manual calibration.
Figure 21 Fragment analysis dye sets example
Manual calibration is recommended—No auto calibration or manual calibration has
1
been performed.
Manual calibration is not required—A manual calibration has been performed.
2
Manual calibration is not required—The factory calibration has been optimized for the
3
instrument by auto calibration.
Manual calibration is recommended—Auto calibration has been performed, but only
4
2 capillaries passed.
Manual calibration is not required—A manual calibration and an auto calibration have
5
been performed.
SeqStudio
Genetic Analyzer Instrument and Software User Guide