Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer User Manual page 207

Hide thumbs Also See for Applied Biosystems SeqStudio Genetic Analyzer:
Table of Contents

Advertisement

SeqStudio
Genetic Analyzer Instrument and Software User Guide
3. Touch Detail to see allele-specific details for fragment/HID analysis and basecall
accuracy for sequencing.
4. Touch Export report (see "Export the install run report" on page 208).
Fragment analysis install run pass/fail criteria
# Allele peaks
# Size standard peaks
Standard deviation of the observed allele fragment sizes
for all 4 capillaries
Appendix C Dye calibration and install standard checks
Result
Perform an install run
Acceptance criteria
15 for all four capillaries
34 for all four capillaries
<0.15 base pairs (bp)
C
207