Recommended Maintenance Schedule

Recommended Maintenance Schedule
Table 9-1.
Routine maintenance tasks
9-2
The following chart shows routine maintenance tasks, the equipment on which
you perform them, and how often you perform each task. The term months refers
to calendar months.
Frequency
Equipment
When used
Handpiece
When used
Tip torquing set
When used
CUSA console
When used
Cooling water
reservoir
When used
Footswitch
Every month or 20
CUSA handpiece
procedures, which-
ever comes first
Every 3 months, or
CUSA console
every 12 to 15
procedures, which-
ever comes first
Every 6 months or
CUSA console
when the surgeon
notices a lack of
suction
Every 6 months or
CUSA handpiece
100 procedures,
whichever comes
first
CUSA EXcel Ultrasonic Surgical System Service Manual
Task
Clean and wipe dry.
Clean and wipe dry.
Clean and wipe dry.
Discard water from the reservoir.
1. Discard the footswitch cover.
2. Clean the footswitch.
Lubricate the handpiece connector
o-rings.
1. Fill the cooling water reservoir with
a solution consisting of 100 cc of
70% alcohol and 900 cc of water.
2. Connect the handpiece, then turn on
the system.
3. Allow the system to run a few
cooling cycles to clean the interior
tubing.
4. Turn off the system.
5. Discard the alcohol solution, then
fill the reservoir with distilled
water.
6. Turn on the system, and allow it to
cycle the water to clean the alcohol
solution out of the tubing.
7. Turn off the system.
8. Discard water from the reservoir.
Replace the contamination guard. Write
either the installation date or the
expiration date on the new guard.
Replace the handpiece connector
o-rings.