Maintain the Tip Torquing Set

Maintain the Tip Torquing Set
9-6

Clean the Tip Torquing Set

Notice
Product damage will result if you do not follow these notices when cleaning the Tip
Torquing Set.
• Do not use ultrasonic or automatic washers.
• Do not autoclave.
• Do not use chlorinated substances such as bleach solution.
• Do not clean with abrasives such as Ajax, Comet, or steel wool.
Items Needed
Detergent, two soft cloths, and rinse water.
Procedure:
Using a soft cloth and a water and detergent combination, manually clean the
1.
tip torquing base and the torque wrench.
Rinse the tip torquing set thoroughly with water.
2.
Using a soft cloth, dry the tip torquing set.
3.

Check the Torque on the Tip Torquing Wrench

A biomedical engineer or an Integra service representative should check the
torque on the tip torquing wrench.
Tools Needed
Torque indicating wrench
7/32 in. hex socket
9/32 in. hex socket.
Procedure:
Attach the socket to the torque indicating wrench.
1.
Clamp the torque indicating wrench handle in a vise.
2.
On the torque indicating wrench, adjust the scale under the indicator needle to
3.
zero.
Set the peak indicating needle on the torque indicating wrench to zero.
4.
Place the green end of the torque wrench from the tip torquing set on the 9/32
5.
in. hex socket.
Rotate the tip torquing wrench clockwise until you feel and hear a click.
6.
Rotate the tip torquing wrench clockwise until you feel and hear a second
7.
click.
CUSA EXcel Ultrasonic Surgical System Service Manual