Install The Control Panel Assembly; Control Head Components - Integra Cusa Excel Service Manual

Ultrasonic surgical aspirator
Hide thumbs Also See for Cusa Excel:
Table of Contents

Advertisement

Important
While the screws are still loose,
take care not to let the control panel
assembly fall out of the control
head.

Control Head Components

CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual

Install the Control Panel Assembly

Remove ESD strap.
1.
Notice
To avoid product damage, remove the ESD wrist strap before connecting cables to the
Machine Control PCB.
Position the control panel assembly at the control head opening.
2.
Before sliding the control panel assembly into the control head, connect the
3.
following cables (Refer to Figure 11-6):
40 pin ribbon cable (J3)
20 pin ribbon cable (J5)
10 pin cable (J4)
6 pin cable (J2).
Slide the control panel assembly into the control head.
4.
Replace four screws (M4 Phillips) and four lock washers (M4), located at the
5.
rear of the control head, that fasten the control panel assembly to the control
head.
Swivel the control head so that the front of the control panel aligns with the
6.
front of the console.
Plug the power cord connector into the wall receptacle.
7.
Turn on the AC Main Switch, located on the rear panel.
8.
The control head contains three replaceable components:
Machine control PCB
Control panel overlay
Display PCB.
The control panel overlay covers the front of the control panel (its fasteners are
accessible inside the control head). The Machine Control PCB and display PCB
(attached to a chassis) constitute two layers in a control panel assembly within the
control head (refer to Figure 11-8). Gaining access to any of these components
involves removing previous layer(s).
Control Head Components
11-15