Open The Left Cover; Open The Right Cover; Open The Rear Cover - Integra Cusa Excel Service Manual

Ultrasonic surgical aspirator
Hide thumbs Also See for Cusa Excel:
Table of Contents

Advertisement

Console Covers
Important
The handpiece cooling water lines
attach to the right cover. The
cooling water lines include a
service loop that allows you to
remove the cover for servicing the
system. Be careful not to remove
the cover far enough to damage
the lines.
11-4

Open the Left Cover

Remove the left side lock screw (M5 Phillips) located under the console top
1.
near the left rear corner.
Push the left side latch toward the front of the console. Tilt the cover outward,
2.
then pull up to remove the cover.

Open the Right Cover

Remove the right side lock screw (M5 Phillips) located under the console top
1.
near the right rear corner.
Remove the cooling water reservoir from its slot.
2.
Push the right side latch toward the front of the console. Tilt the cover
3.
outward, then pull up to remove the cover.
The cooling water lines include a service loop that allows you to remove the
cover for servicing the system. Be careful not to remove the cover far enough
to damage the lines.

Open the Rear Cover

Disconnect the footswitch connector from the receptacle on the rear panel.
1.
Remove two screws (M5 Phillips) from the bottom of the rear cover.
2.
Push the rear cover upwards to unhook it from the console, then remove the
3.
cover.
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual