Replacing The Power Connector - Integra Cusa Excel Service Manual

Ultrasonic surgical aspirator
Hide thumbs Also See for Cusa Excel:
Table of Contents

Advertisement

Table 3-1.
Wire colors
Important
Ensure that all electrical
installations comply with local
codes and regulatory
requirements.
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual

Replacing the Power Connector

If your facility's power receptacles are not compatible with the standard connector
supplied on the CUSA EXcel system, install a different power connector
according to your institution's policy.
Tools needed:
Slotted blade screwdriver
Procedure:
Remove the connector provided with the unit.
1.
Attach the new connector, using the following chart:
2.
Wire Color
green with yellow stripe
brown
light blue
The power connector meets all requirements for safe grounding. Do not defeat its
purpose by using extension cords or three-prong-to-two-prong adapters.
Periodically check the power cord for damaged insulation or connectors. Always
grasp cords by the connector; do not pull on the cord to remove the connector
from the wall receptacle.
If required by local regulations, connect an equipotential grounding cable to the
equipotential grounding lug on the rear of the CUSA EXcel console.
Notice
To avoid product damage, be sure to insert fuses that are appropriate to the power supply:
• 95 to 120 V, T10A, Time Delay (default)
• 220 to 240 V, T5A, Time Delay
Preparing the CUSA EXcel Console for Use
Prong
safety ground
line/hot
neutral
3-9