Figure 11-19.
Removing the suction hose and clamp
from the suction control valve
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Remove 3/8 in. ID suction tube and clamp from the suction control valve: Set
6.
them aside for re-use.
Note: The new suction control valve does not include this tube and clamp.
Install the Suction Control Valve
Assemble the Suction Control Valve to the Manifold Bracket
Attach the 3/8 in. ID suction tube and clamp to the suction control valve.
1.
Attach the suction control valve to the manifold bracket:
2.
Position the suction control valve inlet screen on the manifold bracket so
a.
that its three holes align with the three screw holes in the bracket.
Position the suction control valve on the manifold bracket so that its three
b.
holes align with the three screw holes in the inlet screen and bracket.
Replace three screws (M4 socket head cap screws) in the suction control
c.
valve.
Note: A lock washer secures the screw on each end of the assembly, but no
lock washer secures the third screw.
Suction Control Valve
11-27