CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Notice
If you plan to connect the handpiece for activation, ensure that an I.V. set is available to
supply cooling irrigation solution to the handpiece tip. Otherwise, damage to the tip and/
or handpiece may result.
Electrostatic Sensitive Devices This unit contains electrostatic sensitive components.
When repairing the system, work at a static control workstation. Wear a grounding strap
when handling electrostatic sensitive components. Handle circuit boards by their
nonconductive edges. Use an antistatic container for transport of electrostatic sensitive
components and circuit boards.
When installing a new power supply, use the same screws and washers removed from the
previous power supply. A too-long screw will damage the power supply.
To avoid product damage, do not remove the jumper cable or the connector at J4.
To avoid product damage, remove the ESD wrist strap before connecting cables to the
Machine Control PCB.
To avoid product damage, when handling the suction pump controller card, do not turn
the three white adjustment dials designated as current limit (R13), max speed (R12), and
min speed (R11).

Calibration

Warning
High Current Hazard Do not touch the ultrasonics PCB heat sink while the
CUSA EXcel system is energized. The heat sink is isolated from ground, and can deliver
very high current, resulting in injury.

Repair Policy

Caution
To prevent the ramp from falling and causing injury, be sure to hold the ramp in the up
position while stapling the corrugated tab that secures it to the crate.
Warnings, Cautions, and Notices
1-5