CUSA EXcel Ultrasonic Surgical System Service Manual

Lubricate the Handpiece Connector O-Rings

Keeping the o-rings lubricated prevents them from drying out and cracking,
which may cause handpiece cooling problems.
Items needed: A high viscosity lubricant such as silicone grease.
Items to be lubricated: The four o-rings on the handpiece connector water
prongs.

Replace the Handpiece Connector O-Rings

Regularly replacing the o-rings prevents excessive wear, which may result in
handpiece cooling problems.
Items needed: O-Ring maintenance kit (Part Number 260-650-000), which
includes o-rings, an insertion tool, and instructions.
Items to be replaced: The four o-rings on the handpiece connector water prongs.
Refer to Figure 9-2.
Handpiece connector

Recalibrate the Handpiece

Handpieces for the CUSA EXcel system do not need recalibration unless a
problem develops. If this happens, contact Integra to arrange returning the
handpiece for service. (Refer to Return Equipment for Service in Section 13,
Repair Policy.)
Maintain the Handpiece
Insertion tool
O-ring
9-5