Suction Pump and Controller Card
Table 11-6.
The suction pump controller card
cable to the ultrasonics PCB
Important
The cables in the cable clamps
sometimes tangle. When you pull
the controller card cable away
from the console, be careful to
avoid pulling other cables with it.
Figure 11-26.
Removing the cable clamps and
screws from the suction pump
controller card
11-36

Remove the Suction Pump Controller Card

Disconnect the controller card from the ultrasonics board at J15:
1.
Connector
4-pin
Remove the two cable clamps on the bottom of the controller card bracket,
2.
then separate the controller card cable from other cables in the clamps.
Remove the suction pump controller card from the motor control bracket by
3.
removing four screws (M4 Phillips), four lock washers (M4), and four flat
washers (M4).