CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Status Indicators
To change from Standby to Run, or from Run to Standby, press the Status button.
A green LED illuminates to show when the system is in Run status.
Run:
An amber LED illuminates to show when the system is in Standby
Standby:
status.
An amber LED illuminates in the following circumstances:
Wait:
•
Flashing – Immediately after turning on ( | ) the System Power Switch, the
system automatically fills the handpiece with cooling water
•
Flashing – Immediately after turning off (
system automatically drains handpiece cooling water
•
Continuous – When any alarm condition exists
•
Continuous – When in Prime mode.
An amber LED illuminates in the following circumstances:
Cooling Water:
•
Immediately after turning on ( | ) the System Power Switch, the system
automatically fills the handpiece with cooling water.
•
Immediately after turning off (
automatically drains handpiece cooling water.
Setpoint Adjustment Buttons
Marked with Up/Down arrows, these buttons allow you to adjust the setpoint
values:
10 increments; linear scale of 10 to 100%; default value is
Aspiration (Suction):
100%.
10 increments; linear scale of 1 to 10 cc/min; default value is 3þcc/min.
Irrigation:
10 increments; linear scale of 10 to 100%; default value is 10%.
Amplitude:
5 increments from std (standard operation) to ++++. TissueSelect
TISSUESelect:
interrupts continuous vibration for specified times:
Setting
Std
+
++
+++
++++
*1 ms = 1 millisecond = 1 one-thousandth of a second
Push a button once for a single change in setpoint value. Holding down the button
changes a setpoint value by only one setting (you get no additional setpoint value
changes by holding down the button).
) the System Power Switch, the
) the System Power Switch, the system
On-Time
Off-Time
continuous
0 ms*
40 ms
10 ms
30 ms
10 ms
20 ms
10 ms
10 ms
10 ms
Control Panel
4-7