Standards and IEC Classifications

CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
ATTENTION
Consult accompanying documents.
DANGER
Explosion risk if used with flammable anesthetics.
EQUIPOTENTIAL GROUND
Connect equipotential ground cable here.
POWER ON
Applies power at the AC Main Switch and the System Power Switch.
POWER OFF
Removes power at the AC Main Switch.
PROTECTIVE EARTH GROUND
CAUTION
To reduce the risk of electric shock, do not remove the cover. Refer servicing to
qualified service personnel.
Classified with respect to electrical shock, fire, mechanical, and other specified
hazards only in accordance with UL60601-1 and CAN/CSA C22.2 No. 601.1.

Class I Equipment (IEC 601-1)

Accessible conductive parts cannot become live in the event of a basic insulation
failure because of the way in which they are connected to the protective earth
conductor.

Type BF Applied Part Equipment (IEC 601-1)

The CUSA EXcel console provides a high degree of protection against electric
shock, particularly regarding allowable leakage currents. It is type BF isolated
(floating) output.

IPX - 8 (IEC 529, UL60601) Footswitch

The CUSA EXcel footswitch includes protection against the effects of continuous
immersion in water.
Standards and IEC Classifications
5-5