CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Ultrasonics and Machine Control Printed Circuit Boards

Install the Ultrasonics PCB

Place the new ultrasonics PCB assembly near the CUSA EXcel console. Insert
1.
the following ultrasonic PCB fan assembly lead wires into the 6-pin fan
connector housing:
Refer to Table 11-2 and Figure 11-4.
Position the ultrasonics PCB assembly in the CUSA EXcel console.
2.
Replace four screws (M5 Phillips) and four lock washers (M5) on the
3.
ultrasonics PCB panel.
On the new ultrasonics PCB assembly, the header at J6 (syphon valve) has a
4.
3.96 mm (0.156 in.) center. The CUSA EXcel console syphon valve lead
wires terminate in a 2-pin connector with either a 3.96 mm (0.156 in.) or a
2.54 mm (0.100 in.) center.
If the connector for the console syphon valve has a 3.96 mm (0.156 in.)
center, connect it to the ultrasonics PCB assembly at J6.
OR
ONLY IF the connector for the console syphon valve has a 2.54 mm
(0.100 in.) center, plug the syphon valve cable assembly adapter into the
connector, then connect the adapter/connector assembly to the ultrasonics
PCB assembly at J6. Refer to Figure 11-5.
Adapter for syphon
valve cable
assembly
Syphon valve connector
J6
J6 Detail
11-11