Important
When the screws are loose, take
care not to let the control panel
assembly fall out of the control
head.
Figure 11-9.
Disconnecting control panel cables
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Remove the Control Panel Assembly
Turn off the AC Main Switch, located on the rear panel.
1.
Unplug the power cord connector from the wall receptacle.
2.
Swivel the control head so that the front of the control panel is 90° from the
3.
front of the console.
Remove four screws and four lock washers (M4 Phillips), located at the rear
4.
of the control head, that fasten the control panel assembly to the control head.
Slide the control panel forward far enough to allow disconnecting the
5.
following cables (refer to Figure 11-9):
40 pin ribbon cable (J3)
20 pin ribbon cable (J5)
10 pin cable (J4)
6 pin cable (J2).
Remove the control panel assembly from the control head.
6.
J1
J2
Control Head Components
J3
J5
J4
11-17