Integra Cusa Excel Service Manual page 120

Ultrasonic surgical aspirator
Hide thumbs Also See for Cusa Excel:
Table of Contents

Advertisement

Condition
Excessive misting at the handpiece
tip—continued
No green LEDs (other than setpoints)
illuminate when you activate vibration
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Causes
Suction connections at the suction
canister not seated properly
Open port(s) on the suction canister
Suction canister improperly connected
Inside of the contamination guard wet and
blocked
Problem at the CUSA EXcel console
suction port
Suction pinch valve working improperly
System in Standby mode
User Troubleshooting
Recommended Action
1. Remove any unused suction canister
"elbow" connectors.
2. Reseat all connections.
Close any open ports on the suction
canister.
Refer to the canister manufacturer's
instructions for connecting the canister.
1. Disconnect the contamination guard
fitting from the suction canister
tubing. If you feel no suction at the
end of the contamination guard
fitting, the guard is blocked.
2. Replace the contamination guard.
Remove the contamination guard and put
a finger in the suction system port. If you
feel no suction, call Integra for assistance.
Check the Lap mode indicator.
• If Lap mode is off, in Run mode the
pinch valve should be open.
• If Lap mode is on, in Run mode the
pinch valve should be open only when
you activate vibration.
If these conditions do not exist, call
Integra for assistance.
Press the Status button to switch the
system to Run mode.
10-9