Irrigation Pump Head And Gear Motor; Remove The Irrigation Pump Bezel; Remove The Irrigation Pump Head - Integra Cusa Excel Service Manual

Ultrasonic surgical aspirator
Hide thumbs Also See for Cusa Excel:
Table of Contents

Advertisement

Irrigation Pump Head and Gear Motor

Figure 11-29.
Removing the irrigation pump bezel
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Accessing the irrigation pump head requires removing the irrigation pump bezel;
accessing the irrigation pump gear motor requires removing the bezel and the
panel behind it. The following procedures describe how to remove and re-install
the irrigation pump bezel, head, and gear motor.
Tools Needed
You will need the following tools:
Phillips head screwdriver.
Slotted blade screwdriver
7 mm nut driver
3/32 in. Hex key or Allen wrench.

Remove the Irrigation Pump Bezel

Turn off the AC Main Switch, located on the rear panel.
1.
Unplug the power cord connector from the wall receptacle.
2.
Open the irrigation pump latch half way so that the latch lever stands at the 12
3.
o'clock position.
Four ball-end retainers hold the irrigation pump bezel in place. Remove the
4.
pump bezel by pulling it straight forward. Refer to Figure 11-29.
Irrigation pump bezel

Remove the Irrigation Pump Head

Irrigation Pump Head and Gear Motor
11-41

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents