Maintenance And Care - Integra MAYFIELD 2 Instruction Manual

Skull clamp
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Bleach Dilution:
Caution: Bleach solutions produce chlorine gas. Prepare in a well ventilated area.
NOTE: Hypochlorite potency decreases as bleach ages. Use only recently purchased bleach to prepare
decontamination solutions. Do not use pre-diluted bleach. Prepare a fresh batch for each application.
Hypochlorite bleach mixing guidelines to obtain 20,000 ppm solution:
Concentration
NOTE: The use of highly alkaline agents may cause some components to tarnish and/or corrode.
While this does not affect the performance of the device we recommend that you use an acidic
neutralizing agent immediately after each use as this will help keep your device looking like new.

Maintenance and Care

The MAYFIELD 2 Skull Clamp should be returned to its storage area after each use whether it be a
designated OR or equipment room.
NOTE: Ensure all components are completely dried after each cleaning.
If the Skull Clamp is dropped or mishandled, it should be inspected for damage.
(REF Inspection section of this Instruction manual). If damage occurs, do NOT use;
return the complete device immediately to Integra LifeSciences for inspection.
NOTE: Follow hospital procedures for disposal of this device.
Bleach
4.5%
Pour 4 units bleach into 5 units water
5%
Pour 2 units bleach into 3 units water
6%
Pour 1 unit bleach into 2 units water
Table 5 Hypochlorite Bleach Mixing Guidelines.
CAUTION!
Preparation Instructions
CAUTION!
Page 17

Advertisement

Table of Contents
loading

Table of Contents