Condition
Little or no vibration or fragmentation
(Amplitude LEDs may illuminate well
below setpoint)—continued
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Causes
Tip loose due to improper assembly
User Troubleshooting
Recommended Action
1. Press the Status button to place the
system in Standby mode.
2. Press the Test button to test the
handpiece and tip.
3. If the Vibration alert activates, do one
of the following:
If tip assembled in a nonsterile area:
a. Remove the handpiece from the
sterile field.
b. Use the tip torquing base and
wrench to tighten the tip.
c. Test the tip.
d. Sterilize the handpiece.
If sterile tip assembled to handpiece on
the sterile field:
a. Use the sterilizable torque base
and a disposable sterile wrench to
tighten the tip. (Refer to the
Supplement, Assembling or
Changing Tips on the Sterile
Field, in the CUSA EXcel
Ultrasonic Surgical Aspirator
System User's Guide.)
b. Test the tip.
10-11