Table 9-2.
Torque values for the torque wrench

Store the CUSA EXcel System and Accessories

CUSA EXcel Ultrasonic Surgical System Service Manual
Read and record the peak torque.
8.
Repeat steps three through eight for a total of three times.
9.
Repeat the process on the blue end of the tip torquing wrench and the 7/32 in.
10.
hex socket.
All values must be in the following ranges:
11.
Handpiece frequency
23 kHz (green)
36 kHz (blue)
If the values are higher or lower, send the wrench to Integra for recalibration.

CUSA Console

Drain all liquids. Store the CUSA system in a low traffic area that is free of dirt,
blood, water, and other contaminants. Store the system at an ambient temperature
between -34° C (-29° F) and 65° C (149° F).

Handpiece

Drain all liquids. Store the handpiece in the sterilizer case according to your
facility's policy.
Store the handpiece at an ambient temperature between -34° C (-29° F) and 65° C
(149° F).

Footswitch

If you have attached the footswitch to the console, you do not need to disconnect
it except for maintenance or service. When you are not using the CUSA system,
store the footswitch in the storage compartment in the rear of the console.

CEM Nosecone

The CEM nosecone is a sterile, single-use item. Store unopened packages; discard
used CEM nosecones.
Store the CUSA EXcel System and Accessories
Range (inch-pounds)
55 to 70
20 to 32
9-7