O-Rings On The Cooling Water Reservoir Connectors - Integra Cusa Excel Service Manual

Ultrasonic surgical aspirator
Hide thumbs Also See for Cusa Excel:
Table of Contents

Advertisement

Figure 11-35.
O-rings on cooling water reservoir
connectors
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Integra offers a cooling water preventive maintenance kit. This kit allows you to
perform two tasks:
Replace the cooling water pump tubing.
Replace the o-rings on the cooling water reservoir connectors.
Perform each task every 12 months.
This section tells you how to replace the o-rings on the cooling water reservoir
connectors; the previous section, Cooling Water Pump Tubing, tells you how to
replace the cooling water pump tubing.

O-Rings on the Cooling Water Reservoir Connectors

This replacement requires removing the cooling water reservoir from the console,
then removing the o-rings on the reservoir connectors. The following procedure
describes how to remove the reservoir, remove the used o-rings, install the new o-
rings, and re-install the reservoir.
Press the cooling water reservoir release button (just above the reservoir) on
1.
the arm housing to release the fittings inside the housing.
Slide the reservoir toward the rear of the console.
2.
Remove the o-ring on each male connector on the cooling water reservoir.
3.
Refer to Figure 11-35.
Open (tear or cut the end) the packet of silicone lubricant.
4.
Apply the silicone lubricant to the two o-rings in the cooling water preventive
5.
maintenance kit.
Slide one o-ring over the end and into the groove at the distal end of each male
6.
connector on the cooling water reservoir. Refer to
Figure 11-35.
Re-install the reservoir: Slide it into the slot on the console side panel until it
7.
snaps into place.
Cooling Water Reservoir O-Rings