Install the Suction Pump Controller Card

Notice
To avoid product damage, when handling the suction pump controller card, do not turn
the three white adjustment dials designated as current limit (R13), max speed (R12), and
min speed (R11).
Attach the new suction pump controller card to the motor control bracket
1.
using four screws (M4 Phillips), four lock washers (M4), and four flat
washers (M4).
Replace the two cable clamps on the bottom of the controller card bracket.
2.
Connect the controller card cables:
3.
Loc
Connector
J15
4-pin
(ultrasonics
PCB)
Cable from
7-pin, black
suction
pump
Cable from
3-pin, white
suction
pump
*
The jumper wire occupies pins 6 and 7 on the connector.
Tip: To align the white, 3-pin cable connectors easily, line up the
corresponding segments on the white label that you cut when you removed the
new suction pump from its packaging.
Calibrate the suction pump and controller card. For instructions, refer to
4.
Suction Pump in Section 12, Calibration, in this manual.
Return the arm and control head to normal position.
5.
Replace the left, right, and rear covers.
6.
Plug the power cord connector into the wall receptacle.
7.
Turn on the AC Main Switch, located on the rear panel.
8.
Suction Pump and Controller Card
Wire Color
red, black, orange, brown
yellow, brown, green, red, orange
red, blue, yellow
*
11-37