System 24 Volt Power Supply

System 24 Volt Power Supply
11-6
Replacing the system 24 volt power supply involves removing the left side cover,
disconnecting wiring harnesses, and removing the power supply chassis.
Installing a new power supply reverses the process.
Tools Needed
Phillips screwdriver
7 mm nut driver.

Remove the System 24 Volt Power Supply

Turn off the AC Main Switch, located on the rear panel.
1.
Unplug the power cord connector from the wall receptacle.
2.
Remove the left side cover.
3.
Disconnect the wiring harness (refer to Figure 11-2).
4.
Remove two hex nuts (M4), two lock washers, and two flat washers to release
5.
the power supply chassis from the console.
Slide the power supply and its chassis out the left side of the console.
6.
Remove the power supply and its chassis from the console.
7.

Install the System 24 Volt Power Supply

Replace the power supply in the console.
1.
Slide the power supply and its chassis into the console.
2.
Replace the two hex nuts (M4), two lock washers, and two flat washers to
3.
fasten the power supply to the console.
Connect the wiring harness. Refer to Figure 11-2.
4.
Replace the left side cover.
5.
Plug the power cord connector into a wall receptacle.
6.
Turn on the AC Main Switch, located on the rear panel.
7.
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual