Important
When the screws are loose, take
care not to let the control panel
assembly fall out of the control
head.
Figure 11-6.
Disconnecting control panel assembly
cables
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Ultrasonics and Machine Control Printed Circuit Boards
Remove the Control Panel Assembly
Swivel the control head so that the front of the control panel is 90° from the
1.
front of the console.
Remove four screws (M4 Phillips) and four lock washers (M4), located at the
2.
rear of the control head, that fasten the control panel assembly to the control
head.
Slide the control panel forward far enough to allow disconnecting the
3.
following cables:
40 pin ribbon cable (J3)
20 pin ribbon cable (J5)
10 pin cable (J4)
6 pin cable (J2).
Refer to Figure 11-6.
J1
J2
Remove the control panel assembly from the control head.
4.
J3
J5
J4
11-13