Integra Cusa Excel Service Manual page 102

Ultrasonic surgical aspirator
Hide thumbs Also See for Cusa Excel:
Table of Contents

Advertisement

CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Wait for the Cooling Water and flashing Wait indicators (amber) to turn off
7.
(this indicates that the handpiece cooling water circuit is working properly). It
usually takes about one minute.
Caution
Do not allow the handpiece tip to contact anyone or anything during tip activation.
Contact may result in patient injury, user injury, or handpiece tip damage.
Push the Test button (orange
8.
box on a black button at the
top of the amplitude display
column) on the control
panel.
The system automatically
activates the handpiece at 100% vibration for 4 seconds. During the test, green
LEDs in the amplitude display light up to 100% to show that the handpiece is
working correctly.
Verify the following:
9.
The Vibration alert (red) is off.
The LEDs light up to 100%.
If you see no Vibration alert, and if the LEDs light up to 100%, the handpiece
is working properly.
If the Vibration alert illuminates, or if the LEDs do not light up to 100%, the
handpiece is not working properly. This usually results from an improperly
attached tip. Troubleshoot by
Removing and re-attaching the tip, then pushing the Test button again.
If the problem persists, attach a different handpiece.
If the problem still persists, contact the Integra Service Center nearest you.
If the test is successful, turn off the System Power Switch.
10.
Wait for the handpiece cooling water to drain (the amber Wait and Cooling
11.
Water indicators go off), then disconnect the handpiece.
Handpiece Test (Optional)
8-5