Integra Cusa Excel Service Manual page 125

Ultrasonic surgical aspirator
Hide thumbs Also See for Cusa Excel:
Table of Contents

Advertisement

User Troubleshooting
Condition
Low suction at the handpiece
tip—continued
The handpiece gets hot
The nosecone is loose
10-14
Causes
A blockage in the suction tubing between
the handpiece and the suction canister
Suction canister full
Suction connections at the suction
canister not seated properly
Open port(s) on the suction canister
Suction canister improperly connected
Inside of the contamination guard wet,
blocked, or both
Problem at the CUSA EXcel console
suction port
Suction pinch valve working improperly
Surgeon holding handpiece at tip and flue
Handpiece damage due to tip torquing
without the torque base
Large o-ring not installed correctly, or
missing
Nosecone worn
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Recommended Action
1. Examine the suction tubing along its
entire length for a blockage.
2. Squeeze the blocked area to loosen it.
3. Immerse the handpiece tip in sterile
irrigation fluid briefly. This may clear
the blockage.
Replace the suction canister.
1. Remove any unused suction canister
"elbow" connectors.
2. Reseat all connections.
Close any open ports on the suction
canister.
Refer to the canister manufacturer's
instructions for connecting the canister.
1. Disconnect the contamination guard
fitting from the suction canister
tubing. If you feel no suction at the
end of the contamination guard
fitting, the guard is blocked.
2. Replace the contamination guard.
Remove the contamination guard and put
a finger in the suction system port. If you
feel no suction, call Integra for assistance.
Check the Lap mode indicator.
• If Lap mode is off, in Run mode the
pinch valve should be open.
• If Lap mode is on, in Run mode the
pinch valve should be open only
when you activate vibration.
If these conditions do not exist, call
Integra for assistance.
Hold the handpiece at the nosecone.
Replace the handpiece.
Install the large o-ring in the proper
position on the handpiece.
Replace the nosecone.