Clean the CUSA EXcel Console
Clean the Sterilizer Case
Maintain the Handpiece
CUSA EXcel Ultrasonic Surgical System Service Manual
Frequency
Equipment
Once a year
Tip torquing
wrench
Once a year
CUSA console
You cannot sterilize the CUSA console, but you can clean it, disinfect it, or both.
Use a mild cleaning solution or disinfectant and a damp cloth.
Warning
Electric Shock Hazard Always unplug the CUSA system before cleaning.
Notice
Do not rub, press, or touch any panels with solvents; caustic, corrosive, or abrasive
cleaning or disinfectant compounds; or other materials that could scratch the panels. Do
not use a betadine-based solution; it will cause discoloration.
Do not allow fluids to enter the chassis.
Unplug the unit power cord from the wall receptacle.
1.
Using standard procedures for your facility, thoroughly wipe all surfaces and
2.
cords with a cleaning solution and disinfectant.
Make sure that the surface is completely dry before using the system again.
3.
Use a neutral detergent to clean the sterilizer case.
Notice
Do not clean the sterilizer case with abrasives. Product damage will result.
Maintaining the handpiece involves cleaning the handpiece, lubricating the
handpiece connector o-rings with a high viscosity lubricant such as silicone
grease, or replacing the handpiece connector o-rings. If a problem develops, you
may also need to have the handpiece recalibrated.
Clean the CUSA EXcel Console
Task
Check the torque.
Contact an Integra service
representative to replace the cooling
water pump tubing.
9-3