Advertisement

Environment

Environment
5-4
Warning
Explosion Hazard Do not use the CUSA EXcel system in the presence of flammable
anesthetics or any potentially explosive or flammable atmosphere.

Low Frequency Leakage

Enclosure leakage
current
Patient leakage current

Duty Cycle

Under maximum loading conditions, the CUSA EXcel console is suitable for
ultrasonics activation times of 10 minutes on, 5 minutes off.
Operating temperature
range
Operating humidity
range
Operating atmospheric
pressure range
Storage and shipping
temperature
Storage and shipping
humidity
The system requires a minimum of one hour exposure at its operating temperature
range before you use it.
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
<300 µA
<100 µA
10° C (50° F) to 40° C (104° F)
15 to 80%, relative humidity, non-condensing
70 kPa to 106 kPa (10.15 to 15.38 psi)
-34° C (-29° F) to 65° C (149° F)
25 to 85% relative humidity, non-condensing