Summary of Contents for Integra CUSA Clarity

  • Page 1 Integra CUSA Clarity ® ® Ultrasonic Surgical Aspirator System Cleaning Instruction Guide...
  • Page 2: Intended Audience

    Integra CUSA Clarity – Ultrasonic Surgical Aspirator System ® ® Disassembling, Cleaning, and Sterilizing the System PREFACE Intended Audience This cleaning instruction guide and the equipment it describes are for use only by qualified medical professionals trained in the par- ticular technique and surgical procedure to be performed.
  • Page 3: Disassembling The System

    Disassembling the System CUSA Clarity Items to Keep and Discard After a Surgical Procedure ® Keep Discard Handpiece Nosecone Contamination Guard Flue EUTs (with less than 5 uses) EUTs (after 5 uses) Torque Base Tip Stylet Footswitch Torque Wrench CUSA Quick Connect ™...
  • Page 4 Integra CUSA Clarity – Ultrasonic Surgical Aspirator System ® ® Disassembling, Cleaning, and Sterilizing the System 5. Slide the flue off of the nosecone base along the tip. 6. Twist off the nosecone counter-clockwise until it unlocks. 8. Place the handpiece on the torque base as shown and make 7.
  • Page 5: Disassembling The Console

    Disassembling the Console 1. Detach the aspiration tubing from the patient port on the 2. Disconnect the contamination guard tubing canister. from the vacuum port of the canister. The contamination guard should not be removed NOTE from the console unless it is time to replace it (every six months or when the color changes).
  • Page 6: Cleaning The System

    - Always unplug WARNING CAUTION the CUSA Clarity System before cleaning it. instructions described in the cleaning section. Cleaning the Touchscreen Make sure the console is unplugged before cleaning it. Clean the touch screen with a lint-free wipe and a cleaner, such as window/ glass cleaner, eyeglass cleaner, or a quaternary ammonia disinfectant.
  • Page 7 Cleaning the Handpiece, EUT, and Torque Base The handpiece, extended use tips, and torque base may be cleaned by either the manual cleaning or autowash cycle process prior to sterilization. For Sterlization Parameter, see “Sterilizing the System” on page 8 . Manual Cleaning 1.
  • Page 8 The handpiece and components need to be packaged for sterilization. Integra provides a sterilization tray for steam sterilization of the CUSA Clarity handpiece. This tray protects the handpiece during ster- ilization and during transfer to the sterile field. The handpiece and components must be cleaned prior to sterilization, see “Cleaning the Handpiece, EUT, and Torque Base”...
  • Page 9 The handpiece and components can be packaged in the sterilization tray either: • Wrapped—Sterilization tray double wrapped in hospital CSR material • Flash (Unwrapped)—Sterilization tray unwrapped The cycles below are acceptable for sterilizing the CUSA Clarity handpiece and components: Packaging Temperature...
  • Page 10 Sterilization of Handpieces and Accessories You must sterilize the CUSA Clarity handpiece with steam before use. Some of the CUSA Clarity system accessories are sterile, single-use items. Other accessories are reusable. You must sterilize all reusable accessories with steam before use.
  • Page 12 Ireland Integra, the Integra logo, and CUSA are registered trademarks of Integra LifeSciences Corporation or its subsidiaries in the United States and/or other countries. CUSA Quick Connect is a trademark of Integra LifeSciences Corporation or its subsidiaries. Sani-Cloth is a registered trademark of Professional Disposables International, Inc. neodisher is a trademark of Chemische Fabrik Dr.