Valve Adjustment: Packaged Valve Mode With An Implanted Valve; Preventive Maintenance, Cleaning, And Disinfection - Integra Codman VPV System Manual

Table of Contents

Advertisement

EN – ENGLISH

Valve Adjustment: Packaged Valve Mode With an Implanted Valve

To adjust a newly implanted valve, use a sterile drape to protect the patient from contact with the transmitter. Follow
the instructions below. Note: Select the Packaged Valve mode when a drape is used because the Implanted Valve mode
will likely result in a "REPEAT ADJUSTMENT" message due to the interference caused by the drape.
1.
Follow Steps 1 through 9 of Valve Adjustment: Implanted Valve Mode, with the following exceptions.
a.
select the packaged valve mode
b.
use a sterile drape between patient and transmitter
c.
DO NOT USE ultrasound gel
2.
Follow Steps 7 through 9 of Valve Adjustment: Packaged Valve Mode to complete the adjustment.
3.
Use an x-ray to confirm the valve setting.

Preventive Maintenance, Cleaning, and Disinfection

WARNING: Do not sterilize the program unit or the transmitter, or immerse them in any liquid. Do not allow liquid to
enter the transmitter.
With general care and handling, the program unit and transmitter are virtually maintenance free. The user must do the
following:
1.
Check the equipment covers and interconnecting leads before each use. If damage is present, do not use; return
the equipment for service (see Service and Repair).
2.
After each use and before storing, disconnect the program unit from the power supply and clean the program
unit, the transmitter feet, and the transmitter center rod with a damp cloth and mild detergent. Dry with a clean
cloth.
To disinfect, wipe down the entire system and cables with 70% isopropyl alcohol and a clean cloth.
If the equipment does not operate, return the system for service (see Service and Repair).
6

Advertisement

Table of Contents
loading

Table of Contents